Sony UP-DF550: stripes on film prints always in the same places

What happened

The medical film printer produced stripes on the film that always appeared in the same places on every print and did not disappear after routine cleaning and maintenance.

What was checked

Cause

A factory defect of the thermal print head: defective heating elements simply did not print in those positions, so the marks were identical on every film. Cleaning could not help because there was nothing dirty.

What was done

The thermal print head was replaced. Although the printer was out of warranty, the fault was recognised as a factory defect and the head was covered by the manufacturer; the engineers reported the head alone at about 210,000 RUB in this case. Replacement and setup were done as a procedure from the manufacturer's service documentation by authorised personnel. The outcome was a clean print with no stripes.

Safety

Before any cleaning or work on the print head, unplug the printer from the mains and let the head cool down — it reaches high temperatures and is easily damaged by static discharge.

Lesson: Stripes sitting in exactly the same place on every print point to the print head itself rather than dirt — clean once to rule contamination out, then escalate instead of repeating maintenance.