СпектрАп · Арфа-Ц

Spektrap Arfa-Ts: tube filament does not light, OBR fault at power-up

What happened

When the unit is switched on and kV is selected, it reports fault OBR. The tube filament does not glow, and the upper yellow LED on the filament board stays lit.

What was checked

Cause

The engineers found an open-circuit (burned-out) small-focus filament in the X-ray tube. The generator was correctly reporting no filament current.

What was done

The unit was returned to service after replacement of the X-ray tube, as reported in the case; after replacement it was functionally checked.

Safety

Before disconnecting high-voltage cables or working at the tube sockets, switch off and lock out power and follow the manufacturer's discharge/waiting procedure — the HV generator and cables can hold a lethal charge.

Lesson: When a unit reports no filament and the tube stays dark, a simple continuity check on both focal spots is the fastest way to separate a tube fault from a filament-board fault before ordering parts.

Covered in the course: X-ray Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.