Spektrap Arfa-Ts: tube filament does not light, OBR fault at power-up
What happened
When the unit is switched on and kV is selected, it reports fault OBR. The tube filament does not glow, and the upper yellow LED on the filament board stays lit.
What was checked
- With the unit powered down and the high-voltage safety procedure observed, check the continuity of both focal-spot filaments with a multimeter at the tube socket / cable ends.
- Compare the small-focus reading with the large-focus reading: an open circuit on one filament while the other is intact points to the tube, not to the filament board.
Cause
The engineers found an open-circuit (burned-out) small-focus filament in the X-ray tube. The generator was correctly reporting no filament current.
What was done
The unit was returned to service after replacement of the X-ray tube, as reported in the case; after replacement it was functionally checked.
Safety
Before disconnecting high-voltage cables or working at the tube sockets, switch off and lock out power and follow the manufacturer's discharge/waiting procedure — the HV generator and cables can hold a lethal charge.
Lesson: When a unit reports no filament and the tube stays dark, a simple continuity check on both focal spots is the fastest way to separate a tube fault from a filament-board fault before ordering parts.
Covered in the course: X-ray Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.