Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Strato 2000
Strato · 2000

Strato 2000: table overshoots its travel limit and hits the pedestal

What happened

During an examination the table kept driving into the pedestal — it no longer stopped at its end-of-travel position.

What was checked

Cause

The engineers found that a split (cotter) pin on the table lift gearbox had turned out of its proper position, so the mechanism no longer triggered the optical end-stop sensor in time. The pin had worked loose because of an earlier improvised, non-standard rework of the mechanism.

What was done

The engineers corrected the turned split pin on the gearbox and returned the mechanism to its standard configuration (a new pin of the correct type is safer than reusing a deformed one), then confirmed that the table stops at its end positions. Outcome: solved.

Safety

Before removing covers, isolate the table from the mains and make sure the tabletop cannot move under its own weight; after the repair, test both travel end stops under power before returning the table to clinical use, since a heavy table that ignores its limit is a crush and pinch hazard for patient and staff.

Lesson: On a travel-limit fault, do not stop at the sensor and its wiring — check the mechanical linkage too, because a displaced pin or a past makeshift repair can keep the sensor from ever seeing the end position.

Covered in the course: X-ray Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes