Swelab Alfa: aspirates more blood than the set volume
What happened
As reported in the case: the analyzer draws a larger blood volume than the set sample volume, while measurement results stay correct and no error codes appear.
What was checked
- Check the tubing from the sampling probe to the pump for kinks, leaks or partial blockage.
- Check pump operation.
- With the front cover removed, watch whether blood actually reaches the sample-presence sensor or passes it without being detected.
- Check the sample sensor sensitivity setting in the menu, reported in the case as 0.0.
Cause
Root cause not confirmed. The reported suspicion: sample sensor sensitivity set too low, so blood was not registered in time and aspiration continued past the set volume; tubing or pump problems were not ruled out either.
The cause is probable and was not fully confirmed.
What was done
As reported in the case, the engineers changed the sample sensor sensitivity in the menu from 0.0 to 1.0 and recommended verifying tubing patency and pump operation. The outcome was probable; the root cause was not confirmed.
Safety
The sample path carries patient blood, so work with blood-borne-pathogen precautions and gloves and disinfect any spills, and since the front cover must come off, isolate the analyzer from the mains first and complete the electrical safety check before returning it to use.
Lesson: Correct measurement results do not prove the sampling path is healthy — when the aspirated volume drifts, check the sample-presence sensor and its settings along with tubing and the pump.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.