Swelab · Alfa

Swelab Alfa: aspirates more blood than the set volume

What happened

As reported in the case: the analyzer draws a larger blood volume than the set sample volume, while measurement results stay correct and no error codes appear.

What was checked

Cause

Root cause not confirmed. The reported suspicion: sample sensor sensitivity set too low, so blood was not registered in time and aspiration continued past the set volume; tubing or pump problems were not ruled out either.

The cause is probable and was not fully confirmed.

What was done

As reported in the case, the engineers changed the sample sensor sensitivity in the menu from 0.0 to 1.0 and recommended verifying tubing patency and pump operation. The outcome was probable; the root cause was not confirmed.

Safety

The sample path carries patient blood, so work with blood-borne-pathogen precautions and gloves and disinfect any spills, and since the front cover must come off, isolate the analyzer from the mains first and complete the electrical safety check before returning it to use.

Lesson: Correct measurement results do not prove the sampling path is healthy — when the aspirated volume drifts, check the sample-presence sensor and its settings along with tubing and the pump.