Swelab Alfa: recurring faults traced to separator gel in the rotary valve
What happened
The analyzer misbehaved for a long time with no clear cause; the fault only became evident when the rotary valve was disassembled and gel deposits were found inside.
What was checked
- Disassembly and inspection of the rotary valve, where deposits of separator gel were found.
- Review of operator practice: the laboratory was using tubes with separation gel.
Cause
The engineers found that separator gel from the tubes used by operators had entered the fluidics and contaminated the rotary valve.
The cause is probable and was not fully confirmed.
What was done
Mechanical cleaning of the valve, as reported in the case; flushing of the fluidic path, including with alcohol, was discussed as an additional step. The outcome was reported as probable, not fully confirmed.
Safety
Fluidic parts of a hematology analyzer can hold blood residue: decontaminate the fluid path before disassembly, wear gloves and eye protection, and complete required cleaning and QC before returning the analyzer to use.
Lesson: When a hematology analyzer misbehaves without a clear cause, ask what tubes the operators use — separator gel can slowly contaminate the rotary valve and mimic unrelated faults.