Sysmex CA-600: correct coagulation results only after repeated runs
What happened
As reported in the case: the analyzer kept missing coagulation, and a correct result came out only on the third or fourth attempt, with no error codes shown.
What was checked
- The engineers reviewed which runs dropped their results to see the pattern of the fault.
- They inspected the reagent dispensing path and the cuvette catcher and found reagent occasionally going past the cuvette well instead of into it.
Cause
Reagent was now and then dispensed next to the well, so the cuvette was partly or wrongly filled; a contaminated catcher and a dirty measurement block made the optical reading unreliable.
What was done
The catcher was fully disassembled, serviced and reassembled, and the measurement block was cleaned through. After that the analyzer, as reported in the case, produced a valid result on the first attempt.
Safety
Before opening the fluidics and measurement area, power down and unplug the analyzer, handle all residual samples and reagents as biohazardous with gloves and eye protection, dry the cleaned parts fully, and run the electrical safety test before return to use after wet cleaning.
Lesson: When a coagulation analyzer gives good results only after repeated attempts and shows no error codes, first verify that reagent actually reaches the cuvette before suspecting optics or electronics.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →