Sysmex CA-600: intermittent sampling errors with serum present in every tube

What happened

The analyzer intermittently reported sampling errors even though serum was present in all tubes at the same level.

What was checked

Cause

The engineers did not finally confirm the root cause. They suspected a combination of contamination and needle position drift in the sampling unit with a defective grounding cable on the needle block, whose resistance changed during movement and could disturb serum detection.

The cause is probable and was not fully confirmed.

What was done

Recommended in this case: clean and lubricate the guide rails, check and restore the grounding cable on the needle block, recalibrate the needle and the detection sensitivity, and perform counter maintenance for the sampling unit as a procedure from the manufacturer's service documentation, done by authorised personnel. The outcome was probable, not confirmed by a follow-up observation.

Safety

Treat the needle, tubes and any spilled serum as infectious biohazard — the probe stays sharp even with power off — and repeat the manufacturer's electrical safety checks, including protective earth continuity, before returning the analyzer to use.

Lesson: When sampling errors appear with identical sample levels, test cables under movement, not only statically, and treat cleaning, calibration and counter maintenance as one job.