Sysmex Hemacount 5D: laser line clogging linked to lysing reagent

What happened

The analyzer had problems the engineers traced to the lysing reagent path, with a suspected blockage building up in the tubing on the laser line.

What was checked

Cause

The engineers found sediment accumulating in the tubing of the mixing block that feeds the laser chamber, gradually narrowing the line until it clogged. This was the probable cause reported in the case.

The cause is probable and was not fully confirmed.

What was done

The engineers flushed the entire laser line with a bleach (hypochlorite) solution and replaced the lysing reagent with fresh stock. The outcome was reported as probable, not fully confirmed.

Safety

Hypochlorite is corrosive — wear gloves and eye protection, never mix it with other cleaning chemicals, and remember the fluid path carries potentially infectious material, so flush thoroughly and confirm normal operation before returning the analyzer to use.

Lesson: Stale or off-spec lysing reagent can quietly crystallise in narrow tubing, so check reagent freshness and flush the optical line before blaming the optics or electronics.