Sysmex Hemacount 5D: laser line clogging linked to lysing reagent
What happened
The analyzer had problems the engineers traced to the lysing reagent path, with a suspected blockage building up in the tubing on the laser line.
What was checked
- Inspected the mixing block under the laser assembly, behind the rotary valve.
- Checked the small peristaltic-style tubing runs for sediment buildup and clogging.
- Questioned the lysing reagent itself — freshness and quality — since a poor reagent leaves deposits.
Cause
The engineers found sediment accumulating in the tubing of the mixing block that feeds the laser chamber, gradually narrowing the line until it clogged. This was the probable cause reported in the case.
The cause is probable and was not fully confirmed.
What was done
The engineers flushed the entire laser line with a bleach (hypochlorite) solution and replaced the lysing reagent with fresh stock. The outcome was reported as probable, not fully confirmed.
Safety
Hypochlorite is corrosive — wear gloves and eye protection, never mix it with other cleaning chemicals, and remember the fluid path carries potentially infectious material, so flush thoroughly and confirm normal operation before returning the analyzer to use.
Lesson: Stale or off-spec lysing reagent can quietly crystallise in narrow tubing, so check reagent freshness and flush the optical line before blaming the optics or electronics.