Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Sysmex micro cc
Sysmex · micro cc

Sysmex Micro CC: whistling instead of crackle during aperture burn

What happened

During the aperture burn the engineers heard a whistling sound instead of the usual crackling and suspected a faulty aperture.

What was checked

Cause

In this case the engineers judged the aperture itself to be fine; the probable cause was in the dilution system — valves not working correctly, most likely because the diluent had spoiled (bacterial growth), so air passed through the cell and gave the whistle. The conclusion was probable, not confirmed.

The cause is probable and was not fully confirmed.

What was done

The path recommended in the case: first run the analyzer on saline to confirm the reagent, then check and service the dilution-system valves. A strong warning from the same case: on this model the aperture burn is risky and often ends with the whole aperture chamber needing replacement, so do not use burn cycles as a diagnostic tool; note also that the burn does not start automatically after a batch of samples on this analyzer.

Safety

Treat every fluid you drain from the analyzer as potentially infectious, and power it down and isolate it before working on the aperture, since the burn drives a heavy current through the cell.

Lesson: An odd sound during an aperture burn usually points upstream to the fluid path, so verify the reagent and dilution valves before condemning the aperture — and think twice before running burns on this model.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes