Sysmex XL-680: bent aspiration needle leaking through a crack
What happened
Water was pulsing out horizontally from a crack in the aspiration needle, and the needle was visibly bent. No error codes were reported in the case.
What was checked
- The engineers inspected the aspiration needle and found a bend with a crack, through which water pulsed during operation.
- They asked when the deformation appeared, to rule out damage already present at acceptance; the needle had been straight initially.
Cause
In this single case the cause was reported as most likely mechanical damage — someone reached into the sampler while the analyzer was running; the engineers mentioned a similar case with the same assembly on this model. One case, not a rule.
What was done
The fix was a temporary sleeve repair: a PTFE tube, or a wider hollow needle, fitted and sealed over the crack — soldering the needle material itself is difficult. A donor needle was later found on site, leaving two working needles; the sleeved one should still be replaced with a proper part when available.
Safety
Never reach into the sampler while the analyzer is running — the needle assembly moves without warning and any residue is patient material, so let the cycle finish, shut down per the manual, and wear gloves when handling the needle.
Lesson: A bent aspiration needle almost always means mechanical interference — establish when the damage appeared, keep hands out of a running sampler, and treat any sleeved repair as temporary until a proper needle is fitted.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.