Sysmex · XL-680

Sysmex XL-680: bent aspiration needle leaking through a crack

What happened

Water was pulsing out horizontally from a crack in the aspiration needle, and the needle was visibly bent. No error codes were reported in the case.

What was checked

Cause

In this single case the cause was reported as most likely mechanical damage — someone reached into the sampler while the analyzer was running; the engineers mentioned a similar case with the same assembly on this model. One case, not a rule.

What was done

The fix was a temporary sleeve repair: a PTFE tube, or a wider hollow needle, fitted and sealed over the crack — soldering the needle material itself is difficult. A donor needle was later found on site, leaving two working needles; the sleeved one should still be replaced with a proper part when available.

Safety

Never reach into the sampler while the analyzer is running — the needle assembly moves without warning and any residue is patient material, so let the cycle finish, shut down per the manual, and wear gloves when handling the needle.

Lesson: A bent aspiration needle almost always means mechanical interference — establish when the damage appeared, keep hands out of a running sampler, and treat any sleeved repair as temporary until a proper needle is fitted.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.