Sysmex XN 550: intermittent hemoglobin error with results still correct
What happened
On a hematology analyzer, a hemoglobin (SLS channel) error appeared intermittently, yet the HGB results continued to be issued and looked correct. The error came and went instead of staying active.
What was checked
- Checked the SLS reagent first — its condition and supply gave no reason to suspect the hemolyzing chemistry.
- Inspected the hemoglobin measuring cell; nothing obviously wrong was found there.
- Discussed re-adjusting the blank/sensitivity settings, but held off: with correct results, recalibration would have treated the wrong problem.
- Suspecting the sample path, the engineers found a partial clog in the aspiration needle with small clots caught in it.
Cause
A partially clogged aspiration needle: small clots occasionally disturbed the hemoglobin measurement enough to trigger the error, while most aspirations stayed within limits — so results kept coming out correct.
What was done
The engineers flushed the needle with a syringe and cleared the clots; as reported in the case, the HGB error no longer recurred after that. A manual syringe flush of the aspiration needle was enough — no settings were changed.
Safety
Treat all fluid flushed from the sample probe as potentially infectious: wear gloves and eye protection while flushing, capture the waste in a proper container, and complete the recommended rinse/decontamination cycle plus QC before returning the analyzer to use.
Lesson: When a fault is intermittent but the results stay correct, suspect a partial mechanical blockage in the sample path first — reagent, optics, and calibration settings come after the needle.