Sysmex XN 550: intermittent hemoglobin error with results still correct

What happened

On a hematology analyzer, a hemoglobin (SLS channel) error appeared intermittently, yet the HGB results continued to be issued and looked correct. The error came and went instead of staying active.

What was checked

Cause

A partially clogged aspiration needle: small clots occasionally disturbed the hemoglobin measurement enough to trigger the error, while most aspirations stayed within limits — so results kept coming out correct.

What was done

The engineers flushed the needle with a syringe and cleared the clots; as reported in the case, the HGB error no longer recurred after that. A manual syringe flush of the aspiration needle was enough — no settings were changed.

Safety

Treat all fluid flushed from the sample probe as potentially infectious: wear gloves and eye protection while flushing, capture the waste in a proper container, and complete the recommended rinse/decontamination cycle plus QC before returning the analyzer to use.

Lesson: When a fault is intermittent but the results stay correct, suspect a partial mechanical blockage in the sample path first — reagent, optics, and calibration settings come after the needle.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →