Sysmex XP-300: blank measurement error with vacuum fluctuating under load

What happened

On startup the analyzer failed the blank measurement with an error. The vacuum reading at rest was within range, but the vacuum was suspected to fluctuate during actual operation. Earlier attempts — cleaning the compressor and adjusting the vacuum — had not helped.

What was checked

Cause

The needle of the vacuum regulator was contaminated. At rest the regulator still held the set vacuum, but during operation the dirty needle let the vacuum fluctuate enough to fail the blank measurement.

What was done

The regulator needle was cleaned. Reported in the case as solved: after cleaning, startup and the blank measurement completed without the error.

Safety

Isolate the analyzer from the mains and release residual vacuum and pressure before opening the compressor or regulator, and treat all aspirated samples and wetted parts as infectious.

Lesson: A vacuum reading that looks normal at rest does not clear the pneumatic path — when the fault appears only under load, check the regulator internals for contamination.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.