Sysmex XP-300: blank measurement error with vacuum fluctuating under load
What happened
On startup the analyzer failed the blank measurement with an error. The vacuum reading at rest was within range, but the vacuum was suspected to fluctuate during actual operation. Earlier attempts — cleaning the compressor and adjusting the vacuum — had not helped.
What was checked
- Checked the vacuum value at rest — it stayed within range, so the resting reading alone did not reveal the fault.
- Adjusted the vacuum setting — no lasting improvement.
- Opened and cleaned the compressor, checked it for wear — nothing decisive found.
- Inspected and cleaned the needle in the vacuum regulator — it came out contaminated, which pointed to the actual cause.
Cause
The needle of the vacuum regulator was contaminated. At rest the regulator still held the set vacuum, but during operation the dirty needle let the vacuum fluctuate enough to fail the blank measurement.
What was done
The regulator needle was cleaned. Reported in the case as solved: after cleaning, startup and the blank measurement completed without the error.
Safety
Isolate the analyzer from the mains and release residual vacuum and pressure before opening the compressor or regulator, and treat all aspirated samples and wetted parts as infectious.
Lesson: A vacuum reading that looks normal at rest does not clear the pneumatic path — when the fault appears only under load, check the regulator internals for contamination.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →