Sysmex XS-1000i: both RAID hard drives failed
What happened
The control PC of the hematology analyzer started failing: both IDE drives of the RAID array (80 GB each, as reported in the case) died one after the other, and a first attempt to substitute SATA drives through IDE adapters did not work.
What was checked
- Localise the fault: determine at which step or subsystem the failure shows up (boot, analysis run, data access).
- Check each drive of the array separately — in this case both were found dead, so the RAID had no surviving disk left.
- Reinstall the Control Side and User Side software from scratch and restore the database from the .mdb backup.
Cause
Failure of both IDE drives of the RAID array — the engineers found the disks had simply worn out and died one after the other, leaving the system without any working drive.
What was done
New IDE drives of the original type were sourced (officially expensive, as noted in the case); the Control Side and User Side software was reinstalled and the database restored from the .mdb backup — this was a procedure from the manufacturer's service documentation, done by authorised personnel. The SATA-through-adapter substitution did not work. The case notes also mention similar units fixed by reseating boards, re-connecting ribbon cables and tightening board screws.
Safety
Before returning the analyzer to sample testing, complete the electrical safety test and follow biohazard precautions when handling aspirated blood and fluid waste.
Lesson: On older analyzers the hard drives are often the first components to die, so keep the database backup current and line up proven replacement drives before the RAID loses its second disk.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.