Thermo Indiko: unstable dispensing with air in the sample line

What happened

The analyzer dispensed inconsistently from cycle to cycle, and air was seen entering the aspiration line; the PTFE tube from the sampler was later found cracked.

What was checked

Cause

The engineers found a split in the PTFE tube between the sampler and the dosing block; air was drawn in with each aspiration, so the dispensed volume varied from cycle to cycle.

The cause is probable and was not fully confirmed.

What was done

Replace the damaged tube with the correct spare part and check the line for air after the replacement. In this case the reported outcome was probable — confirm stable dispensing with control material before relying on results.

Safety

Treat the sample line and any spilled fluid as biohazardous, keep hands clear of the moving sampler while watching it run, and run quality control before returning the analyzer to service.

Lesson: Unstable dispensing on a chemistry analyzer usually starts in the fluid path — check tubing and fittings for air leaks before blaming electronics or optics.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.