TMO MR-01: Inverter shows zero voltage, unit inoperative

What happened

The unit displayed a message that no voltage was present at the inverter (reported in the case as 0 V) and would not operate.

What was checked

Cause

A mains voltage surge during renovation work in the building had blown the inverter fuse; the fuse did its protective job, and the lost supply produced the zero-voltage indication.

What was done

The engineers replaced the blown fuse with a correct replacement of the same type and rating, after which the unit returned to normal operation (solved in this case).

Safety

Isolate the unit from the mains and prove dead before opening the inverter — its capacitors can hold a dangerous charge — fit only the fuse type and rating from the manufacturer's documentation, and complete electrical safety checks before the unit goes back into clinical use.

Lesson: A zero-voltage inverter message can simply mean a protective fuse did its job after a mains disturbance, so check the supply path and any recent site work before condemning the inverter itself.

Covered in the course: Mammography Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.