Tomey FT-1000: broken joystick stem and sticking two-level joystick
What happened
The metal stem of the two-level tonometer joystick snapped off, and the joystick itself periodically sticks or behaves erratically.
What was checked
- Inspection of the joystick mechanism — in this case the engineers found screws in the lower part of the joystick working loose and jamming the mechanism.
Cause
The sticking came from screws in the lower part of the joystick unscrewing over time and jamming the mechanism; the broken stem was apparently the result of forcing a joystick that would not move freely.
What was done
A new metal stem was made to match the original by a machinist, and the plastic handle was reattached — reported in the case as done either with two-part epoxy ("cold weld") or with a solvent cement made by dissolving a piece of the same plastic in dichloroethane (toluene). The loose screws were the root cause, so they need to be secured, not just the parts reglued. Outcome reported as solved.
Safety
Disconnect the instrument from the mains before opening the joystick assembly, work with dichloroethane- or toluene-based adhesives only with strong ventilation and away from ignition sources, and confirm free movement of the joystick through both levels plus an electrical safety test before the unit returns to patient use.
Lesson: A stiff or erratic control usually has a simple mechanical root — check for loose fasteners and wear before forcing the part, and repair the cause, not just the damage it caused.
Covered in the course: Fault-Finding Method for Medical Equipment →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.