TOPCON · TRK-1P

TOPCON TRK-1P: joystick position not detected at power-up

What happened

On every power-up the autorefractometer asked for joystick calibration and did not register the position of the control lever; the calibration prompt would not clear.

What was checked

Cause

Faulty joystick: the position sensors inside no longer produced a valid signal, so the instrument could not see the lever position and treated it as a unit needing calibration.

What was done

Replace the joystick as a complete assembly. In this case the calibration prompt disappeared right after the replacement and no calibration was needed; the instrument worked normally. Checking the old sensor resistances first is a reasonable way to confirm the diagnosis before ordering the part.

Safety

Unplug the instrument from the mains before removing any covers, and confirm it passes a basic electrical safety check before it is returned to use.

Lesson: A calibration prompt that reappears at every power-up usually points to a worn joystick rather than a software setting — check the sensor resistances first, then replace the assembly.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.