Toshiba 1.5T MRI: patient flipped by magnetic pull of an undeclared knee prosthesis
What happened
During positioning, before scanning started, the patient was suddenly pulled over against the bore wall — a knee endoprosthesis had snapped onto the magnet. As reported, three staff were needed to free her; no injuries.
What was checked
- Standard pre-scan screening about implants, metal, pacemakers and dentures was done — the patient had forgotten the removable prosthesis, so it was not declared.
- Only after the incident did inspection of the leg show a removable knee prosthesis hidden under hosiery, plastic on the outside.
Cause
A removable knee prosthesis with ferromagnetic parts under a plastic cover, hidden by hosiery and forgotten by the patient, was caught by the always-on static field and jerked her towards the bore.
What was done
Three staff pulled the patient clear, as reported, and she was unharmed. After the case the routine changed: patients now change out of their own clothes and are checked before entering the magnet room, so hidden items are found before they reach the magnet.
Safety
The MR static field is always on and cannot be switched off: keep ferromagnetic and unverified items out of the magnet room, and if a patient is pinned, follow the site's MR emergency procedure with enough staff — never pull alone.
Lesson: Verbal screening alone cannot guarantee MRI safety — the patient must change clothes and every hidden item must be found before entering the magnet room.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.