Toshiba 1.5T MRI: patient flipped by magnetic pull of an undeclared knee prosthesis

What happened

During positioning, before scanning started, the patient was suddenly pulled over against the bore wall — a knee endoprosthesis had snapped onto the magnet. As reported, three staff were needed to free her; no injuries.

What was checked

Cause

A removable knee prosthesis with ferromagnetic parts under a plastic cover, hidden by hosiery and forgotten by the patient, was caught by the always-on static field and jerked her towards the bore.

What was done

Three staff pulled the patient clear, as reported, and she was unharmed. After the case the routine changed: patients now change out of their own clothes and are checked before entering the magnet room, so hidden items are found before they reach the magnet.

Safety

The MR static field is always on and cannot be switched off: keep ferromagnetic and unverified items out of the magnet room, and if a patient is pinned, follow the site's MR emergency procedure with enough staff — never pull alone.

Lesson: Verbal screening alone cannot guarantee MRI safety — the patient must change clothes and every hidden item must be found before entering the magnet room.