Toshiba 64-slice CT: reconstruction queue error during heavy back-to-back load

What happened

During long runs of back-to-back reconstructions on a Toshiba 64-slice CT, the reconstruction queue threw an error once more than 3000 images had accumulated in it (figure as reported in the case).

What was checked

Cause

Queue overflow: the engineers found that the pipeline cannot offload images fast enough, and once the queue passed roughly 3000 images (as reported in the case) the system stopped with an error. This was observed in one case, not as a general limit.

What was done

No parts were involved. The working workaround reported in the case: keep the queue below 2000-2500 images (about 3-4 patients scanned in a row), and when the error appears, reboot the system following the manufacturer's shutdown/startup procedure. Colleagues additionally suggested limiting the number of slices per batch to 900-1300 and not chasing throughput records. All figures are as reported in the case, not verified limits.

Safety

After any reboot, let the system complete its full startup self-checks and confirm reconstruction and image integrity are normal before returning the scanner to clinical use.

Lesson: Treat the reconstruction queue as a load buffer with a practical ceiling — pacing batch throughput prevents overflow errors that a reboot only clears temporarily.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.