Toshiba Alexion: fault when writing scan data to the raw data drive
What happened
The scanner threw a fault at power-up or during scanning. In this case the logs showed reconstruction starting normally, and the error appeared only at the moment the system tried to write scan data to the reconstruction data drive (RDD).
What was checked
- Read the logs first: reconstruction started cleanly each time, so the recon chain was not the obvious suspect and the fault pointed to the write path.
- Checked what came before the startup fault: in this case the error acknowledged with OK was usually preceded by another message, which gave a clue to the real source.
Cause
The engineers found that the RDD itself was failing: the drive returned an error while scan data was being written, so the reconstruction side looked healthy in the logs while the storage side was not.
What was done
Re-initializing the RDD, following the procedure from the manufacturer's service documentation, cleared the fault; reported as solved, no parts were replaced.
Safety
Before returning the scanner to clinical use after any drive or system-level intervention, verify the interlocks and radiation warning indicators and complete the manufacturer's electrical safety and exposure checks — never resume scanning if any protection had to be bypassed to keep the system running.
Lesson: When the logs show a stage starting cleanly but the failure hits exactly at the write step, look at the storage subsystem first and try a documented re-initialization before ordering hardware.
Covered in the course: CT Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.