Toshiba · Aplio 300

Toshiba Aplio 300: Image noise on the 3.5 MHz probe only

What happened

Interference artefacts appeared on the ultrasound image only when working with the 3.5 MHz probe; all other probes produced a clean image, and no error codes were shown.

What was checked

Cause

External electromagnetic interference from two LED light fixtures near the exam position. The 3.5 MHz probe turned out to be the most sensitive to it, which made the fault look like a defective probe rather than a room problem.

What was done

The two LED lights identified as the interference source were switched off and removed from the exam area; after that the image with the 3.5 MHz probe was clean. No parts were replaced and no service settings were touched.

Safety

When tracing an interference source, switch equipment off via its normal controls or the wall switch, and leave any mains wiring or fixed-lighting work to an electrician — never open luminaires or junction boxes live.

Lesson: Noise on a single probe can come from the room, not the probe — before condemning a transducer, rule out nearby electrical interference by switching suspect sources off one at a time.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.