Toshiba Aplio 500: incomplete startup after software image reinstall
What happened
After reinstalling the software images, the scanner booted into Windows but startup never completed — the application did not come up fully and the data disk volumes were not accessible.
What was checked
- Check that the drive letters assigned to each partition match the volume labels the software expects, on both the system disk and the data disk.
- Verify the disk and partition layout after imaging before concluding that the images or the drives themselves are faulty.
Cause
The engineers found that after re-imaging, the partition letters no longer matched the volume labels, so the software could not locate its data volumes and startup stalled.
What was done
The partition letters were remapped to match the correct volume labels; after that the scanner completed startup normally. The case was reported as solved.
Safety
After any work on storage or software, complete the electrical safety checks required before return to use and verify normal acquisition on a phantom before the scanner is used on patients.
Lesson: When a system fails to start fully after re-imaging, check that drive letters and volume labels match the expected layout before suspecting the images or the hardware.