Toshiba Aplio: trackball rollers crumbling, cursor unstable
What happened
Cursor driven by the control panel trackball moved erratically - jerky, skipping, sometimes dead. On opening the trackball, the small foam drive rollers were hardened and crumbling, with debris packed around the shafts.
What was checked
- Dismantled and cleaned the trackball; the engineers found the debris came from the disintegrating foam rollers, and cleaning alone gave only brief improvement.
- Measured roller diameter and width to cut like-for-like replacements.
- Compared candidate materials reported in the case: PPU foam, polyethylene foam and foamed silicone.
Cause
PPU foam is the wrong material for these rollers: reported in the case to harden and crumble after 3-6 months of use, while foamed silicone is the material intended there by design.
What was done
The engineers cut new rollers from foamed silicone sheet with a hole punch sized to the original diameter and pierced the centre holes with a ballpoint pen refill; casting rollers from the softest silicone in a homemade mould was offered as an alternative. Case outcome: solved - cursor stable again, and silicone rollers are reported to last far longer than foam.
Safety
Isolate the system from mains before opening the control panel, and complete the electrical safety (leakage current) test before the machine returns to clinical use.
Lesson: When a control degrades again and again, the wear material itself is the fault - replacing crumbling foam with a dimensionally identical, chemically stable one beats any amount of cleaning.