Toshiba · Aquilion 16

Toshiba Aquilion 16: voice commands not recording

What happened

On a Toshiba Aquilion 16 CT, voice commands stopped being captured during recording. The voice command subsystem did not register any input, with no error codes reported in the case.

What was checked

Cause

The sound card drivers had been installed incorrectly, so the voice command subsystem received no audio input. This was the cause found in this single case, not a general rule.

What was done

Reinstalling the sound card drivers restored recording of voice commands. The case was reported as solved, with no parts replaced.

Safety

Voice prompts are part of patient communication during scanning, so before returning the scanner to use, verify that all patient instruction and intercom paths work as intended.

Lesson: Before blaming the application or its configuration, check that the underlying driver layer is healthy — a bad driver install can look exactly like a subsystem failure.