Toshiba · Aquilion 32/64

Toshiba Aquilion 32/64: IDD hard drive replacement

What happened

The hard drive in the CT scanner's data storage unit (IDD) needed replacement, and a compatible HDD had to be sourced and prepared for the storage subsystem.

What was checked

Cause

As reported, the original IDD drive had failed, and the storage subsystem uses a non-standard logical sector format (520 bytes in this case), so an ordinary off-the-shelf drive is not recognised until it is reformatted.

What was done

The engineers bought a new server HDD with the same specifications, reformatted it to the 520-byte logical sector format via an Ubuntu console (as reported), and installed it into the IDD of a working CT. The scanner reportedly accepted the drive and it remains in operation. The drive's own firmware was not modified; whether every similar drive will behave the same way was not confirmed in this case.

Safety

Power the scanner down and isolate it from the mains before opening the IDD, take ESD precautions when handling drives, and let the system complete its startup self-tests plus an electrical safety check before the CT is returned to clinical use.

Lesson: A failed OEM storage drive is not always irreplaceable: a matching standard server HDD, reformatted to the system's native sector layout and thoroughly tested beforehand, can do the job — but validate the drive and the system's acceptance before relying on the swap.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.