Toshiba Aquilion 64: 'over kV' error on the high-voltage generator

What happened

A Toshiba Aquilion 64 CT kept reporting an 'over kV' error, pointing to the high-voltage side of the generator.

What was checked

Cause

A faulty 24 V power supply feeding the high-voltage multiplier was the cause of the 'over kV' error in this case.

What was done

The 24 V power supply feeding the HV multiplier was replaced; the error cleared and the diagnosis was confirmed in this case.

Safety

Before opening the generator cabinet, isolate the CT from the mains and follow the manufacturer's high-voltage discharge procedure — multiplier and generator capacitors can hold a lethal charge long after power-off.

Lesson: An 'over kV' error on a CT does not always mean a failed multiplier or tube — check the simple supply rails feeding the HV stages, such as the 24 V line, first.