Toshiba Aquilion 64: scan protocol selection unavailable after patient registration
What happened
After registering a patient, the console would not open scan protocol selection. Tube warm-up and calibration routines still ran normally.
What was checked
- The engineers found that the reconstruction block was fully loaded — recon and related processes all reported ready.
- They compared which functions worked (warm-up, calibrations) against which failed (protocol selection), which pointed to the console software layer rather than the reconstruction hardware.
Cause
A fault in the console software on this unit (version v3.35gr007, as reported in the case). Hardware checks found nothing wrong.
What was done
The console software was reinstalled — a procedure from the manufacturer's service documentation, done by authorised personnel. Restoring from an earlier saved backup was also an option for this type of failure. After the reinstall, protocol selection worked again. Outcome: solved.
Safety
After any console software reinstallation, let the full system start-up and self-checks complete, repeat the required warm-up and calibrations, and run an electrical safety test before returning the scanner to patient use.
Lesson: When some console functions work and others stay blocked after patient registration, suspect the software layer first — confirm subsystem readiness, then clear the fault with a clean reinstall or a known-good backup.