Toshiba · Aquilion 64

Toshiba Aquilion 64: scan protocol selection unavailable after patient registration

What happened

After registering a patient, the console would not open scan protocol selection. Tube warm-up and calibration routines still ran normally.

What was checked

Cause

A fault in the console software on this unit (version v3.35gr007, as reported in the case). Hardware checks found nothing wrong.

What was done

The console software was reinstalled — a procedure from the manufacturer's service documentation, done by authorised personnel. Restoring from an earlier saved backup was also an option for this type of failure. After the reinstall, protocol selection worked again. Outcome: solved.

Safety

After any console software reinstallation, let the full system start-up and self-checks complete, repeat the required warm-up and calibrations, and run an electrical safety test before returning the scanner to patient use.

Lesson: When some console functions work and others stay blocked after patient registration, suspect the software layer first — confirm subsystem readiness, then clear the fault with a clean reinstall or a known-good backup.