Toshiba · Aquillion

Toshiba Aquillion: console RAID down after a drive failure

What happened

The console's RAID array collapsed after a single hard drive failed, leaving the system down. Per the case, this was not the first RAID failure on this unit.

What was checked

Cause

A hard drive at the end of its service life. Drives in these arrays wear out one by one over time, so on an aging unit failures tend to arrive one after another; the engineers also noted that the controllers and the drive cage themselves can fail from time to time.

What was done

The failed drive was replaced with one from old spares, and per the case the system came back up and worked. Whether non-Toshiba drives are a safe substitute remains an open question — drive firmware differences may matter, so the conservative choice is drives as specified in the manufacturer's service documentation.

Safety

Isolate the console from the mains before swapping drives, and return the scanner to clinical use only after the array has fully rebuilt and the manufacturer's startup self-tests pass.

Lesson: Console RAID drives are wear parts: replace them proactively and keep documented compatible spares, rather than letting a single failed drive take the whole scanner down.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.