Toshiba · Artida

Toshiba Artida: no power-on despite standby voltage present

What happened

The ultrasound system would not start. In this reported case the standby 5 V was present, but the turn-on never happened, and a faulty Schottky diode assembly was found in the first power supply.

What was checked

Cause

Probable: degraded Schottky rectifier assemblies in the power supplies. Since replacing the first assembly alone did not restore power-on, the engineers suspected the second PSU's diodes were also blocking the start sequence; the full chain was not confirmed in this case.

The cause is probable and was not fully confirmed.

What was done

Replace the Schottky diode assemblies in both power supplies; the suspicious diodes in the second unit were also slated for replacement. The outcome is probable — power-on had not been confirmed yet. Whether the first PSU may run with the second disconnected was left open; such a question should be settled against the manufacturer's service documentation, with the work done by authorised personnel.

Safety

Disconnect the system from the mains and let the PSU capacitors fully discharge before touching the boards, and after any power supply repair run an earth-bond and leakage-current test before returning the machine to use.

Lesson: A live standby rail does not mean every power supply is healthy — check all PSUs before blaming the main board.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.