Toshiba/Canon Aquilion: NRA detector calibration order and phantom setup
What happened
An engineer asked how to run the NRA detector calibration on an Aquilion CT correctly — the sequence of calibrations and phantom positioning were unclear.
What was checked
- Run the Air calibrations before starting NRA, and plan to repeat them after NRA.
- Select all tube voltages at once and calibrate the whole table, not individual lines.
- Start NRA with the LL phantom: first slice at the 'neck', then at the middle of the main body.
- On the remaining phantoms, place the slice at the centre in both Z and Y.
- Follow the system's on-screen prompts at every step.
Cause
No hardware fault was involved. Reported in the case as a procedural question: running Air and NRA out of order or with a misplaced phantom would produce unreliable calibration data, so the correct sequence and phantom setup had to be confirmed first.
What was done
The engineers found a working order: complete all Air calibrations first, then run NRA starting with the LL phantom (slice first at the 'neck', then at the middle of the main body), keep the remaining phantoms centred in Z and Y, follow the system prompts throughout, and finish with another set of Air calibrations. Exact steps and parameters follow the manufacturer's service documentation. The engineer confirmed the question was resolved.
Safety
NRA calibration produces a long series of X-ray exposures, so before starting make sure the scan room is clear, the door is closed and staff nearby know the tube will be energised.
Lesson: A calibration is only as good as its sequence and setup, so confirm the whole procedure — Air before and after, exact phantom positioning — before you start rather than repeating a failed run.
Covered in the course: CT Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.