Toshiba/Canon Aquilion Prime SP 160: cannot add Semar and a venous phase to scan protocols
What happened
Protocol setup did not work as expected: the engineers could not add the Semar function to the Femoral Runoff protocol, or a venous phase to the CT angiography protocol for the upper-extremity vessels.
What was checked
- Checked the protocol menu: copied the base protocol, edited the required parameters and saved the copy.
- Confirmed with colleagues that Semar is not an in-protocol switch — it can be applied afterwards to the raw data over a selected short range.
- Confirmed that a venous phase is not something that can be appended to an existing protocol but a separate acquisition with its own contrast injection parameters.
Cause
In this case the engineers found two different reasons behind the difficulty. Semar is a post-processing option: it can be applied after the scan to the raw data over the needed range, so it does not have to be built into the acquisition protocol. The venous phase cannot simply be added — it is a different study with its own injection parameters (reported in the case: 150 ml of contrast medium, 3 ml/s, acquisition starting at the 180th second).
What was done
The setup was done through the protocols menu: copy the base protocol, change the required parameters and save. Semar was applied after the scan, to the raw data, over a short section, so as not to wait for a long reconstruction of extra series. The venous phase was configured as a separate study with its own injection parameters.
Safety
Before the scanner returns to clinical use, verify the edited protocols with a test acquisition (on a phantom where possible) and have the changed contrast injection parameters confirmed by the responsible clinician — protocol mistakes directly affect patient dose and contrast delivery.
Lesson: Distinguish what can be changed inside a scan protocol, what is applied afterwards to the raw data (post-processing such as Semar), and what is a separate acquisition with its own injection parameters (a venous phase) — then console setup follows naturally.