Toshiba claude-haiku-5-5: contrast leaking under lower gantry cover, injector lines blowing off
What happened
In this single case, contrast agent seeped through a gap under the lower gantry cover of a CT scanner, building up drips that hung like stalactites inside. Separately, injector lines tore loose under bolus pressure, spilling contrast on the floor, the machine and occasionally the patient, and ruining the study.
What was checked
- Inspection and flushing of the accessible gantry areas after the leaks were spotted
- Warming the contrast before injection to reduce viscosity
- Saline chase after the contrast bolus to flush the lines
- Checking the lines for manufacturing defects and plastic quality
- Checking the line routing for kinks and correct use of injector pressure limits
- Matching catheter gauge to flow rate (green at 5 mL/s, pink at lower rates, as reported in the case) and ruling out extended lines
- Checking the catheter backflow valve when flushing through the upper port
Cause
No single cause was confirmed. The probable explanation reported in the case: defective or poor-quality lines and components, combined with excessive pressure from kinked or badly routed lines, cold viscous contrast and improper handling of the catheter.
The cause is probable and was not fully confirmed.
What was done
The measures reported in this case, with a probable outcome: warm the contrast to body temperature, run a saline chase after the bolus, secure the line and bottle with tape including to the patient's skin, select the catheter by the set flow rate, never extend the lines, and replace defective line batches. The contrast under the lower cover was cleaned by the service engineer during maintenance, with the covers off.
Safety
Isolate the scanner from mains power before removing covers to clean contrast inside the gantry, check wiring and connectors for residue, and confirm electrical safety before returning the system to use.
Lesson: Recurring injector line blow-offs and contrast under gantry covers point first to consumable quality, line routing and contrast temperature, not to the scanner itself.
Covered in the course: CT Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.