Toshiba CT (16-slice, as reported in the case): radiation leak around the viewing window during scanning
What happened
During scanning, radiation was reported behind the leaded viewing window — the dose was escaping the protection into the control side rather than staying behind the glass.
What was checked
- A radiation survey of the control room and adjacent rooms was performed by a radiation protection specialist while the scanner was operating.
- The engineers searched for documentation on the glass to confirm its stated lead equivalent, but no conclusive paperwork could be found.
Cause
The engineers found that the glass had no documented lead equivalent, and the leak was traced to the perimeter of the glazing — the joint with the frame — so the effective shielding there was insufficient.
What was done
Leaded rubber was laid around the perimeter of the glass, and the measured leak reportedly disappeared. The case was closed as solved, but since the gasket could not be confirmed in place later, the exact cause should be treated as probable rather than fully proven.
Safety
Any intervention on the scanner's shielding, even a gasket around a viewing window, changes the radiation barrier: work only with the responsible radiation protection officer, verify the result with a calibrated survey meter during an actual exposure, and have the shielding formally accepted before the room returns to normal use.
Lesson: A protective barrier is only as strong as its joints, so undocumented shielding components must be treated as unverified until a calibrated radiation survey under real exposure proves otherwise.
Covered in the course: CT Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.