Toshiba Excelart: start-up error in the gradient cabinet, reset does not always take

What happened

At power-up the scanner reports a gradient-related error. It can be cleared with the reset control on the status page, but in this case not always on the first attempt.

What was checked

Cause

In this case the suspected cause was breakdown of tantalum capacitors, 20 µF / 15 V (the engineers reported about eight in a row) in the gradient unit, with the associated 2 A fuse blown. This was a probable cause, not a confirmed one from the case alone.

The cause is probable and was not fully confirmed.

What was done

Replace the failed tantalum capacitors and the blown 2 A fuse in the affected block. Fault isolation was done per the gradient troubleshooting section near the end of the manufacturer's service documentation.

Safety

De-energise and lock out the system, allow the storage capacitors of the gradient amplifier to discharge and verify absence of voltage before working inside the cabinet; repeat the electrical safety tests before the scanner returns to clinical use.

Lesson: When a start-up fault resets but keeps coming back, work from the cabinet's own indicators down to the fuses before changing boards — and treat cooling and humidity as prime suspects in gradient electronics failures.