Toshiba Nemio XG: battery warning at first power-on of the day
What happened
A battery warning shows only at the first power-on of the day; after that the scanner runs normally and no settings are lost. When the case was written up, the backup cell had not been measured yet.
What was checked
- Measure the lithium cell voltage on the system board with the unit powered down; in this case the suspicion was that it had dropped below 3 V.
- Check the pattern: does the message appear only after a long shutdown (overnight) and not after a short restart? That fits a weak backup cell more than a board fault.
- Confirm that date/time and user settings really are held after a full power-off.
Cause
Probable, not confirmed: the CMOS backup lithium cell on the system board has aged, and after a long power-off its voltage sits at the level that triggers the warning on a cold start, while the remaining charge is still enough to hold the settings. In this case the engineers only suspected this — no measurement was taken.
The cause is probable and was not fully confirmed.
What was done
The suggested fix was to replace the CR-type lithium cell with a fresh one and verify that the warning no longer appears at the next cold start. In this case the replacement had not yet been carried out, so the outcome is probable, not verified.
Safety
Disconnect the unit from mains and let internal capacitors discharge before opening it; handle the system board with ESD precautions, and run the electrical safety test before returning the scanner to clinical use.
Lesson: A warning that appears only at the first start-up after a long shutdown points first to a weak backup cell, even when settings still hold — measure the battery before suspecting anything deeper.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.