Toshiba Titan T3: MRI fed from a UPS with no output neutral

What happened

An engineer questioned whether a three-phase UPS (reported in the case as 100 kW) could correctly supply the MRI without a neutral conductor on its output, with the scanner fed through an input transformer.

What was checked

Cause

No component fault was reported. The doubt arose only because the UPS output carries no neutral; the engineers established that the neutral for the MRI load is created downstream of the UPS, not supplied by it.

What was done

The engineers confirmed the scheme as installed is correct: the transformer primary is connected in delta, so the UPS output needs no neutral, and the secondary is connected in star with an artificial neutral point that provides the neutral for the MRI. Comparable arrangements were reported at other manufacturers' installations, so no rewiring was required.

Safety

Treat UPS outputs and transformer terminals as live at all times: isolate, lock off and prove absence of voltage before any work, since the battery-backed stage of a UPS can keep circuits energised even with the mains input open, and verify earthing and neutral bonding per the installation drawings before return to use.

Lesson: Before judging a supply scheme wrong, trace where the neutral is actually created in the chain — a delta-wye transformer often provides it where the UPS does not.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.