Toshiba · Xario 660

Toshiba Xario 660: touch screen needs recalibration

What happened

The touch screen required calibration — touches were registering inaccurately on the control panel.

What was checked

Cause

The engineers found that the cause was simple calibration drift of the touch panel, with no signs of hardware failure reported in the case.

What was done

Calibration was completed successfully with the manufacturer's touch-panel calibration utility supplied with the panel drivers, following the service documentation.

Safety

After any work on the console, confirm normal system operation and perform the standard electrical safety test before returning the unit to clinical use.

Lesson: When a touch screen misregisters input, run the standard calibration routine first — not every touch fault means a failed panel.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.