Ulrich 257 injector: head/stop fault after long idle

What happened

After roughly a year and a half out of service (as reported in the case), the injector kept raising a message to check the swivel head or the mechanical stop, and at the same time showed a low internal battery charge. The message came back every time.

What was checked

Cause

The built-in battery had failed after the long idle period; the injector read the resulting power loss during movement as a mechanical problem, which made the head/stop message misleading.

What was done

The internal battery was replaced; after that the error no longer appeared and the injector worked normally both on mains and on battery. The engineers confirmed the battery as the cause.

Safety

Isolate the injector from mains before opening it for battery work; do not short the removed cell and treat it as hazardous waste per local rules, and complete the electrical safety test before returning the injector to use.

Lesson: After a long idle period, check the internal battery before blaming motors, sensors or mechanics — a dead battery can imitate a motion fault.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.