Ulrich · CT Motion XD8000

Ulrich CT Motion XD8000: battery charge lasts a single injection

What happened

The injector held a full battery charge for only one contrast injection; after that the pack was flat and had to be recharged again.

What was checked

Cause

The two batteries had simply aged and lost most of their capacity, so a full charge was no longer enough for a working session of injections.

What was done

As reported in the case, both batteries were replaced with new ones while the injector stayed powered. No settings were lost because they are kept in the touch control panel; after the swap the injector again worked through injections on one charge (reported as solved).

Safety

Battery terminals can deliver very high short-circuit current: use insulated tools, keep rings and metal objects away from the terminals, and before any hot swap confirm from the manufacturer's documentation that it is allowed for this model — otherwise power the unit down as documented; carry out the electrical safety test before the injector returns to use.

Lesson: Injector batteries are wear parts — when runtime per charge drops sharply, plan a like-for-like replacement and first confirm where the device actually stores its settings.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.