Ultrasonix SonixSP: black desktop after boot, ultrasound application not starting
What happened
Reported in the case: after Windows loads, the screen shows only a black desktop; the ultrasound application does not start automatically.
What was checked
- The engineers confirmed that the application exam.exe was not starting on its own at boot.
- They then ran it manually via Explorer from the folder C:\Program Files\Ultrasonix\Exam.
- If the executable is missing from that folder, the installation should be treated as damaged and reinstalled.
Cause
Not established in this case: why the automatic start of exam.exe stopped working after Windows loaded was not determined, and the case was not carried through to a confirmed resolution.
The cause is probable and was not fully confirmed.
What was done
Probable fix, as reported in the case: start exam.exe manually via Explorer (C: > Program Files > Ultrasonix > Exam). If the file is missing, reinstall the software with the manufacturer's installer — a procedure from the manufacturer's service documentation, done by authorised personnel.
Safety
After any software intervention, do not return the system to clinical use until the electrical safety test has been repeated and normal scanning with the connected probes has been verified.
Lesson: When a device seems dead on boot, first check whether Windows itself is healthy and only the application autostart failed — try starting the application manually before going deeper.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.