Academy / Field fault notes / Ultrasonix SonixSP
Ultrasonix · SonixSP

Ultrasonix SonixSP: black desktop after boot, ultrasound application not starting

What happened

Reported in the case: after Windows loads, the screen shows only a black desktop; the ultrasound application does not start automatically.

What was checked

Cause

Not established in this case: why the automatic start of exam.exe stopped working after Windows loaded was not determined, and the case was not carried through to a confirmed resolution.

The cause is probable and was not fully confirmed.

What was done

Probable fix, as reported in the case: start exam.exe manually via Explorer (C: > Program Files > Ultrasonix > Exam). If the file is missing, reinstall the software with the manufacturer's installer — a procedure from the manufacturer's service documentation, done by authorised personnel.

Safety

After any software intervention, do not return the system to clinical use until the electrical safety test has been repeated and normal scanning with the connected probes has been verified.

Lesson: When a device seems dead on boot, first check whether Windows itself is healthy and only the application autostart failed — try starting the application manually before going deeper.