Urit Urit-5160: WBC flag on the first control runs after start-up

What happened

The analyser took a long time to reach its working state, and the first six to seven control measurements, as reported in the case, produced incorrect results; a star flag kept showing next to the WBC values.

What was checked

Cause

The engineers took this as a probable cause: deposits or contamination built up in the WBC measurement chambers made the impedance and optical readings drift apart, and the analyser flagged the disagreement with a star.

The cause is probable and was not fully confirmed.

What was done

As recommended in the case, the WBC measurement chambers were to be filled with a hypochlorite (bleach) solution and left to stand for about 15 minutes, then rinsed thoroughly before the analyser went back to use. The outcome was reported as probable: the cleaning was expected to remove the deposits and clear the flag, but this was not confirmed.

Safety

Hypochlorite is corrosive and the fluid path may hold blood-derived material, so wear gloves and eye protection, ventilate the area, and rinse the chambers fully and confirm a passing control run before returning the analyser to samples.

Lesson: A WBC flag caused by two-method disagreement, repeating on the first runs after start-up, usually points to dirty measuring chambers, not to a sample problem.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.