Urit Urit-5160: WBC flag on the first control runs after start-up
What happened
The analyser took a long time to reach its working state, and the first six to seven control measurements, as reported in the case, produced incorrect results; a star flag kept showing next to the WBC values.
What was checked
- Read what the optical channel measured in the WBC line: the star is raised when the difference between the two WBC measurements is large, so a wide gap between the impedance and optical results points at the measuring chambers rather than at the sample.
- Repeat control runs one after another to confirm that only the first measurements after start-up are affected.
Cause
The engineers took this as a probable cause: deposits or contamination built up in the WBC measurement chambers made the impedance and optical readings drift apart, and the analyser flagged the disagreement with a star.
The cause is probable and was not fully confirmed.
What was done
As recommended in the case, the WBC measurement chambers were to be filled with a hypochlorite (bleach) solution and left to stand for about 15 minutes, then rinsed thoroughly before the analyser went back to use. The outcome was reported as probable: the cleaning was expected to remove the deposits and clear the flag, but this was not confirmed.
Safety
Hypochlorite is corrosive and the fluid path may hold blood-derived material, so wear gloves and eye protection, ventilate the area, and rinse the chambers fully and confirm a passing control run before returning the analyser to samples.
Lesson: A WBC flag caused by two-method disagreement, repeating on the first runs after start-up, usually points to dirty measuring chambers, not to a sample problem.