Villa Vision (CPI G100RF): fluoroscopy lost after generator battery replacement

What happened

After a discharged backup battery in the CPI G100RF generator was replaced and the settings were re-entered, fluoroscopy stopped working: no high voltage was delivered to the image intensifier during fluoro, and no error messages appeared.

What was checked

Cause

In this case the engineers found that during the reconfiguration the two workstation configurations had been assigned to each other's places; the generator therefore saw the fluoro request as coming from the wrong position and produced no high voltage to the image intensifier — without logging a single error.

What was done

The workstation assignments were set back correctly in the generator configuration — a procedure from the manufacturer's service documentation, done by authorised personnel — and fluoroscopy worked again; no parts were needed. The engineer also shared saved CPI configuration backups with colleagues so a future battery failure takes minutes, not hours.

Safety

Radiation safety: before the unit goes back to use, test fluoroscopy at minimum dose settings behind a protective screen, and confirm the exposure warning indicator lights and the beam cuts off instantly when the exposure switch is released.

Lesson: A generator configuration re-entered by hand after a battery loss can hide small swaps that fail silently, so restore from a backup whenever possible and test every operating mode before calling the job done.

Covered in the course: X-ray Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.