Vital Images Vitrea: CT console cannot send data to the workstation

What happened

The CT console could not see the Vitrea workstation and would not send data to it. AE titles were configured on both sides and the hosts answered ping, yet no transfer went through.

What was checked

Cause

The engineers found the problem in the stored DICOM node configuration itself: old or incorrect records in the system. The visible settings looked correct, but the entries behind them were no longer valid.

What was done

In this case the fix was to delete the old DICOM node records and create new ones. After that the console reached the workstation and data transfer was restored.

Safety

After any change to DICOM routing, confirm with the clinical staff before the system returns to use that studies reach only the intended destination — a wrong node can silently send patient data elsewhere.

Lesson: If hosts ping but DICOM traffic fails, treat the stored node configuration as the suspect — clean, valid entries matter more than settings that merely look correct.

Covered in the course: DICOM, PACS and Networks — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.