Vital Images Vitrea: CT console cannot send data to the workstation
What happened
The CT console could not see the Vitrea workstation and would not send data to it. AE titles were configured on both sides and the hosts answered ping, yet no transfer went through.
What was checked
- Check that the AE titles on both sides match exactly, character for character.
- Ping the hosts in both directions to confirm basic network reachability.
- Run a DICOM echo (verification) test between the console and the workstation to see whether the association is accepted.
- Review the stored DICOM node entries on the workstation for old, duplicate or damaged records.
Cause
The engineers found the problem in the stored DICOM node configuration itself: old or incorrect records in the system. The visible settings looked correct, but the entries behind them were no longer valid.
What was done
In this case the fix was to delete the old DICOM node records and create new ones. After that the console reached the workstation and data transfer was restored.
Safety
After any change to DICOM routing, confirm with the clinical staff before the system returns to use that studies reach only the intended destination — a wrong node can silently send patient data elsewhere.
Lesson: If hosts ping but DICOM traffic fails, treat the stored node configuration as the suspect — clean, valid entries matter more than settings that merely look correct.