Voluson E8: permanent key rejected after software reinstall
What happened
After the software was reinstalled on this unit, the permanent key was no longer accepted and the system would not come up in its normal configuration.
What was checked
- The engineers checked that the unit's serial number was being read correctly by the system.
- They compared the installed software release with the release the key was issued for.
Cause
In this case the engineers found the wrong software version had been loaded: it could not read the unit's serial number, and the keys are bound to a specific software release and option set, so a mismatched release will not accept the key.
What was done
The fix was to install the software release matching the existing key; the software load and key handling were a procedure from the manufacturer's service documentation, done by authorised personnel. With the correct release in place the key was accepted, and the case was reported as solved.
Safety
After any software work on an ultrasound system, confirm normal boot and full function and complete an electrical safety test before returning the unit to clinical use.
Lesson: Before reinstalling software on a key-protected system, record the installed release first and load exactly that release, because the key follows the software version, not the other way around.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.