Vyaire Newmovent Tecme: large leak flagged on circuit test
What happened
During the circuit check the ventilator reported a leak of 10 L/min (as reported in the case), while no leak could be heard at the valve block. A second, healthy unit of the same model showed about 0.7 L in comparison (as reported in the case).
What was checked
- Listened along the valve block for audible leakage — the engineers found nothing audible.
- Compared the reading with a known-good unit of the same model.
- Checked the pressure regulator block, the pneumotachograph for possible fluid ingress, and the check valves.
- Leak-tested the expiratory valve for tightness.
Cause
No cause was confirmed in the case. The engineers found nothing audible at the valve block and nothing suspicious in the regulator block, pneumotachograph or check valves, so the probable cause was an internal leak through the expiratory valve — quiet at the block but large enough to fail the circuit test.
The cause is probable and was not fully confirmed.
What was done
The engineers installed a new expiratory valve. According to the discussion, the repeat leak test result had not yet been reported, so this remains a probable fix, not a confirmed one.
Safety
Before valve work, isolate the ventilator from the mains and from the gas supply, and before return to use run the manufacturer's full leak test, verify all alarms, and complete an electrical safety test.
Lesson: A large reported leak with nothing audible means the loss is internal — compare readings with a healthy unit and work systematically through the pneumatic path, with the expiratory valve among the first suspects.
Covered in the course: Ventilator Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.