Vyaire Vela: large gap between inspired and expired volumes
What happened
On a Vela ventilator, expired volume readings diverged strongly from the inspired volumes. The unit was treated as 'dead' and not worth repairing, though it was never formally written off.
What was checked
- Run a full leak test on the circuit and exhalation path first — a leak alone can produce a large inspired/expired gap.
- Remove and inspect the exhalation valve membrane for wear, deformation or debris.
- Check the exhalation flow sensor itself for drift or damage before condemning it.
Cause
In this single case the engineers traced the discrepancy to a degraded exhalation flow sensor giving false expired-volume readings. The outcome is probable, not a rule: leaks and a faulty valve membrane must be excluded before blaming the sensor.
The cause is probable and was not fully confirmed.
What was done
The exhalation flow sensor was replaced — in this case with a spare from a third-party source, which the engineers say calls for extra verification — followed by calibration. The calibration was a procedure from the manufacturer's service documentation, done by authorised personnel. After the swap and calibration the inspired and expired readings agreed again, though the outcome is reported as probable.
Safety
Work on the exhalation path only with the ventilator disconnected from the patient and from mains power, and do not return it to clinical use until it passes the full leak test and complete self-test with delivered volumes verified against the settings.
Lesson: A large inspired/expired volume gap points first at leaks and the exhalation valve, and a flow sensor replacement — original or not — is only finished after authorised calibration and full pre-use testing.
Covered in the course: Ventilator Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.