Wiznet WizFi210: Wi-Fi module fault on a syringe pump

What happened

A syringe pump stopped communicating over Wi-Fi. The WizFi210 module was found faulty, and the repair required replacing it with board-level soldering, including the antenna connection.

What was checked

Cause

The original Wi-Fi module had failed, and the replacement bought for it was a different hardware version (_ex) that is not a drop-in fit, because it lacks the antenna connection of the original.

What was done

The engineers replaced the failed module by soldering a new WizFi210 onto the board and attached the antenna by soldering it directly to the antenna pad on the module. As reported in the case, the rework took several attempts before the link worked reliably, and the replacement module was sourced from a third-party supplier.

Safety

Disconnect the pump from mains and battery before any soldering, work with ESD precautions around the RF module, and complete the electrical safety (leakage current) test plus a full Wi-Fi function check before returning the pump to clinical use.

Lesson: Before ordering a replacement subassembly, confirm the exact hardware version and its antenna and connector differences — the same base part name does not guarantee a drop-in fit.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.