Woson Tanco: false vacuum at rest and vacuum error in cycle

What happened

As reported in the case: before start, with the door open, the display showed a vacuum of -3.3 instead of the normal 0–0.3; after a cycle was started, pressure climbed to about +0.9, and after 5–7 minutes, around 50°C, a vacuum-building error (E12) appeared.

What was checked

Cause

A small metal tooth from the gear of a dental contra-angle handpiece had entered the vacuum circuit and lodged under a rubber seal, partially restricting the line and creating the false vacuum reading.

What was done

The engineers removed the debris from under the seal and reassembled the circuit; a trial cycle with an empty chamber then ran without the vacuum error, and the unit went back into use after passing that test.

Safety

Chamber water, condensate and residues in a sterilizer count as potentially contaminated biohazard — wear gloves and eye protection when draining and cleaning, and confirm the repair with a full empty-chamber test cycle before the unit handles clinical loads again.

Lesson: Before condemning a vacuum pump, inspect the whole vacuum path for debris — instruments shed small parts that can lodge anywhere in the line and mimic a pump or sensor failure.

Covered in the course: Steriliser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.