Woson Tanzo C23: intermittent vacuum fault in sterilisation cycles

What happened

Sporadic err12: the error shows up only from time to time, with the vacuum phase failing intermittently. In the reported case, two cycles run with the engineer on site passed clean and the vacuum built evenly, so the fault could not be reproduced on demand.

What was checked

Cause

The engineers found that the normally-open steam release valve behind the chamber had gummed up and acted only every other time: when it stuck partly open it bled the chamber during the vacuum phase, so the error came and went. Since the fault could not be shown on site, this is the probable cause in this single case, not a rule.

The cause is probable and was not fully confirmed.

What was done

Disassemble and clean the steam release valve as a temporary measure, then fit a new normally-open valve. As supporting work: clean under the door gasket, and replace the gasket or adjust the door if the seal is questionable.

Safety

Before opening any steam or vacuum lines, confirm the chamber is fully depressurised and cooled down — trapped steam and hot condensate can scald even after a finished cycle.

Lesson: An intermittent vacuum error on an autoclave often points to a sticking valve in the steam path rather than the obvious door seal — if cycles pass clean on site, go after the valves that must operate every single cycle.

Covered in the course: Steriliser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.