Woson · Tanzo E23

Woson Tanzo E23: vacuum pump does not start during warm-up

What happened

As reported in the case: about four minutes after the program started, during the warm-up stage, the sterilizer showed error E12 and the vacuum pump did not start at all.

What was checked

Cause

The engineers found an open circuit in one winding of the vacuum pump motor: the pump could not start, and the unit raised the error at the point in the cycle where vacuum is first required.

What was done

Since an open winding is not repairable in the field, the fix was replacement of the vacuum pump; as reported in the case, after replacement the fault was cleared and the unit completed the program without the error.

Safety

Isolate the sterilizer from the mains before working on the pump and verify the run capacitor is discharged before touching its terminals, and confirm the unit passes a leak test and a full sterilization cycle before return to use.

Lesson: When a pump fails to start at the same point in every cycle, work through the basics in order — capacitor, free rotation, winding resistance — before blaming the electronics.

Covered in the course: Steriliser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.